Nitrogen Oxides Analyzer (NO, NOX, NO2)
Single chamber nitrogen oxide analyzer
The new Chemiluminescent NO-NO2-NOX analyzer, model AC32M, combines our 30 years of experience with its predecessors AC30M & AC31M, with an enhanced electronics package and a modular component parts design
Advantages
- Graphic Liquid Crystal Display (LCD)
- Real-time synoptic flow diagram display
- User programmable ranges and average times
- Temperature and pressure compensation
Description
Specification
Applications
Resources
The outcome is an ultra compact and light- rack 3U, easy-to-use, chemiluminescence based analyzer capable of measuring nitrogen oxides at ppb levels. Applied to nitrogen oxides measurement, Chemiluminescence corresponds to an oxidation of NO molecules by O3 molecules. The return to a fundamental electronic state of the excited NO2 molecules is made by luminous radiation, detected by the PM tube.
- Lower detectable limit (2 sigma): 0.4 ppb
- Response time: automatic and programmable (minimum 30 sec)
- Zero drift: less than 0.5 ppb / 24 h less than 1 ppb / 7 days
- Span drift: less than 0.5 % / 24 h less than 1 % / 7 days
- Linearity: +/- 1 % of F.S.
- Mesure en continu des oxydes d’azote dans l’air ambiant, surveillance du trafic automobile
- Mesure de la pollution de fond et recherche sur les oxydes d’azote (NOx)
- Mesure du NO2 dans les gaz médicaux