Handheld conductivity meter D2
Solution for conductivity mesurement of fuels
The SETA D2 handheld conductivity meter provides an accurate and rapid conductivty measurement of fuels.
Advantages
- Conductivity measurement range : 0 – 2000 pS/m
- Fully temperature compensated measurement
- USB interface for data transfer
- High accuracy, AC measurement technology
The SETA D2 handheld conductivity meter meets the requirements of ASTM D2624. It provides an accurate and rapid conductivity measurement of distillate fuels. It is specifically designed for testing low conductivity fluids such as aviation kerosene.
Jet fuel typically has a low conductivity value (between 0,2 and 50 pS/m) and is classified as a static accumulator which can cause danger in a receiving tank. Static dissipator additive is used to increase the fuels conductivity range and counter balance these problems but the effectiveness of the additive diminishes with the time when transferred along pipelines and whilst in storage facilities. As a result it is important to be able to measure conductivity throughout the distribution system.
The meter can measure fuel electrical conductivities between 0 and 2000 picosiemens/meter (pS/m), although it is optimised and normally used in the 0 to 500 pS/m range.
- Conductivity measurement range : 0 – 2000 pS/m
- Accuracy measurement : +/- 1,5 pS/m
- 2,6 AHr lithium-ion battery
- 316 SS coaxial electrode sensor K=0,2
- Resolution measurement : 0,1 pS/m
- ASTM D1655 et D2624
- IP 274
- DEF STAN 91-91